| Model | Model | Measuring Range | Resolution | Measuring Head Width* | SCHMIDT Calibration Material** |
| ETB-100 | ETX-100 | 0.3 – 100.0 cN | 0.1 cN | 24 mm | PA: 0.20 mm Ø |
| ETB-200 | ETX-200 | 2.0 – 200.0 cN | 0.1 cN | 24 mm | PA: 0.20 mm Ø |
| ETB-500 | ETX-500 | 2.0 – 500.0 cN | 0.1 cN | 24 mm | PA: 0.20 mm Ø |
| ETB-1000 | ETX-1000 | 3 – 1000 cN | 0.1 cN | 38 mm | PA: 0.30 mm Ø |
| ETB-2000 | ETX-2000 | 3 – 2000 cN | 0.1 cN | 38 mm | PA: 0.50 mm Ø |
| Calibration: | According SCHMIDT factory procedure |
| Accuracy: | ±1 % full scale ±1 digit or better (typical ±0.5 % full scale) |
| Measuring units: | cN, g, N or lb, switchable |
| Overrange: | Approx. 10 % full scale, without accuracy guarantee |
| Overload protection: | 200 % full scale |
| Measuring principle: | Strain gauge bridge |
| Measuring roller deflection: | Max. 0.5 mm |
| Signal processing: | Digital |
| Damping: | Adjustable electronic damping, averaging |
| Sampling rate internal: | approx. 1 kHz |
| Sampling rate: | Max. 200 Hz (200 readings/sec) |
| Display: | Colour-TFT 128 x 160 |
| Display update rate: | Approx. 2 times per second |
| Memory: | Last, average, MAX and MIN |
| Temperature coefficient: | Gain: less than ± 0.01% full scale/°C |
| Temperature range: | 10 – 45 ºC |
| Air humidity: | 85 % RH, max. |
| Auto power off: | Approx. after 3 minutes of non-use |
| Power supply: | LiPo accumulator (approx. 20 h continuous use, charging time approx. 3.5 h), USB AC adapter 100…240 V AC with 4 adapters (EU/USA/UK/AUS-NZ) |
| Housing material: | Aluminium |
| Dimensions: | See dimensions |
| Weight, net (gross): | Approx. 310 g (1220 g) |